OPC Multichannel Monitor


 


MAIN CHARACTERISTICS

OPC MULTICHANNEL MONITOR – Main features

  • Measurement principle: laser scattering on elliptical mirror
  • Optical channels: 8 optical channels with thresholds having lower limits [µm]: 0.28, 0.4, 0.5, 0.7, 1.1, 2.0, 3.0, 5.0
  • Sampling type: The instrument gives counts per minute, or counts per liter for each optical channel
  • Dimensional sensitivity: 0.28 µm (50 ± 10% count efficiency)

OPERATING CONFIGURATIONS


  • The system can operate in Standalone configuration or in integrated configuration with SWAM 5a Dual Channel Monitor
  • Standalone Configuration: Periodical manual setting of the parameters for the estimation of the real mass concentration values
  • Configuration integrated with the FAI SWAM 5a Dual Channel Monitor: Automatic correction with real mass values supplied by SWAM DC with self-learning procedure

TYPICAL APPLICATIONS

  • – Characterization of the temporal trends of the particulate matter granulometric distribution as a support and integration to information about air quality
  • Real-time estimation of the PM1, PM2.5, PM10, TPS mass concentration of the airborne particulate matter